Ademhalingsfilters
Hydrofoob filter
Beschrijving
MBS casing with connector for capnometry. This device can be placed near the patient between the endotracheal tube and the Y connector to protect the… Lees meer
Beschrijving
MBS casing with connector for capnometry. This device can be placed near the patient between the endotracheal tube and the Y connector to protect the respiratory circuit of a contaminated patient.
However it can only be used in that position for a short time, because this device does not allow humidification nor heating of ventilation gases. More frequently, the respiratory filter is placed next to the ventilator at the distal end of the expiratory tubing, but other positions are possible.
The filter must be changed at least every 24 hours.
Complies with ISO 9360 requirements. Not sterile.
HEPA type antibacterial and antiviral filter.
HEPA: High Efficiency Particulate Air Filtration. The hydrophobic membrane is not impregnated by fluids which ensure a continuous filtrating functionality. Mechanical type filtration.
Efficiency of the bacterial/viral filtration: 99.9999 / 99.99993 % for 3µm particles and 99.9995 % for 0.3 µm particles. Connections: 15 mm male and 15/22 mm female/male.
Recommended tidal volumes: 250 – 1500 ml.
10/240
Referenties en specificaties
Filter | Verpakking | |||||
---|---|---|---|---|---|---|
Favourites | Code | Weight g | Lenght mm | Priming volume ml | Eenheden/Doos | Eenheden/pakket |
553.03 | 47 | 118 | 78 | 10 | 240 |
Aanvullende informatie
-
Aanwezigheid van latex Nee
-
Aanwezigheid van producten van dierlijke of biologische oorsprong Nee
-
Pyrogeenvrij Nee